Esophageal Doppler Monitor

exp date isn't null, but text field is

Placed by Respiratory therapists much like a Naso-gastric tube

Principles:

  • Ultrasound technology used to give real time information about the left ventricular flow based on waveform analysis of descending aortic pulse.

Direct flow measurement

  • Corrected Flow Time (FTc):
    • Indicates left ventricular preload
    • Normal range is 0.330 – 0.360 seconds
  • Peak Velocity (PV):
    • Used as an index to assess contractility
    • Age dependent normal values
  • Stroke Distance (SD)
  • Minute distance (MD)
  • Mean Acceleration (MA)

Calculated measurements:

  • Cardiac output (CO)
  • Stroke volume (SV)
  • Cardiac Index (CI)

Special Thanks

Attribution

If reusing this content please use the following information to provide credit to the content authors:  

  1. Title: Esophageal Doppler Monitor
  2. Author:  Michigan Medicine
  3. Source: The URL where the image is hosted.
  4. License: “CC BY-NC 4.0”

Esophageal Doppler Monitor,  Michigan Medicine, Dept of Surgery,  “CC BY-NC 4.0”